Service and Maintenance Information
4500 Series of Instruments
System User Guide
126 of 138
RUO-IDV-05-0264-A
4. Empty the drain bottle and then dispose of the waste.
5. Install the cap on the bottle.
6. Tighten the clamps holding the hoses to the cap.
Figure 13-4 Source Exhaust Drain Bottle
Item
Description
1
Connection to mass spectrometer
2
Source exhaust drain bottle (In this drawing, the capped drain bottle is
shown at the back of the mass spectrometer to make connection points
visible. The drain bottle is located at the side of the mass spectrometer
when installed.)
3
Connection to vent
Caution: Asphyxiation Hazard: Take care to vent exhaust gases
properly. The use of mass spectrometers without adequate
ventilation to outside air may constitute a health hazard. In addition,
certain procedures required during the operation of the mass
spectrometer may cause gases to be discharged into the exhaust
stream; under these conditions, inadequate ventilation may result
in serious injury. Regularly perform an inspection of the hoses to
make sure that there are no leaks.
Note:
Source exhaust hose connections at the drain bottle, mass
spectrometer, and the lab vent are secured with a hose clamp.
2
3
1
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...