Operating Instructions — Tune and Calibrate
Run the Verify instrument performance option weekly or after the mass spectrometer is cleaned
to confirm that the system is working properly. In general, for triple quadrupole systems,
calibration and resolution is satisfactory for three to six months unless the system loses vacuum.
For linear ion trap (LIT) systems, the resolution should also be satisfactory for three to six months
but the system should be calibrated approximately monthly. If the system loses vacuum, then
check the calibration and resolution before using the system. For more information about tuning
and calibration, refer to the
Advanced User Guide
and the
Manual Tuning Tutorial
.
Optimize the Instrument
The following procedure describes how to verify the performance of the instrument. For more
information about using the other instrument performance options, refer to the Help.
1. In the navigation bar, under
Tune and Calibrate
, double-click
Manual Tuning
.
2. Run a calibration method, and confirm that the TIC is stable and that the peaks of
interest are present in the spectrum.
3. In the navigation bar, under
Tune and Calibrate
, double-click
Instrument
Optimization
.
The
Instrument Optimization
dialog opens.
The instrument name and the current operating mode configured in the active
hardware profile is shown at the bottom of the dialog.
4. Click
Verify instrument performance
.
5. Click
Next
.
6. Click
Approved Tuning
.
7. Click
Next.
Tip!
Perform maintenance tasks regularly to make sure that the mass spectrometer is
performing optimally. Refer to
.
Required Materials
• Tuning solutions that are supplied in the Standards Chemical Kit shipped with the system. If
needed, a new kit can be ordered from AB SCIEX.
• 5 mL, 1 mL, and 250 µL serial gas-tight syringes.
• red PEEK sample tubing.
Prerequisites
• A printer is configured.
• The spray is stable and the correct tuning solution is being used.
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...