Operating Instructions — Analyze and Process Quantitative Data
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
111 of 138
6. If required, select the
Display Low/High values
check box and then examine the
Low
,
High
, and
Mean
for each row in the grid. Each row represents standards that
have the same concentration levels.
7. Select another analyte.
The results are shown on a per-analyte basis.
8. To check for Quality Control variations at the same concentration levels, select
QC
in
the
Sample Type
field.
Compare Results Between Batches
The number of analytes and the analyte names must be the same for the data to be combined in
the Statistics pane.
1. Open the Results Tables.
2. Click
Tools > Statistics
.
3. Do one of the following:
•
To arrange the results by Results Table, in the
Conc. as Rows
list, select
Group By Batch
.
•
To arrange the results in order of concentration, in the
Conc. as Rows
list,
select
Group By Concentration
.
•
To arrange the results in order of concentration, but without a row showing the
statistics for each group or batch, in the
Conc. as Rows
list, select
Group By
Concentration (no All)
.
The software sorts the results. At the end of each batch or group, one or two
additional rows are shown:
All
(statistics for all results tables in that group) and
Average
(statistics on the statistics for that batch or group).
Table 11-1 Integration Tab and Quantitation Wizard Icons
Icon
Name
Function
Set parameters from
Background Region
Uses the selected peak.
Select Peak
Uses the selected background.
Manual Integration Mode
Manually integrates peaks.
Show or Hide Parameters
Toggles the peak-finding parameters between
shown and hidden.
Show Active Graph
Shows the analyte chromatogram only.
Show Both Analyte and IS
Shows the analyte and its associated
chromatogram (available only when an associated
internal standard exists).
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...