
Operating Instructions — Batches
4500 Series of Instruments
System User Guide
64 of 138
RUO-IDV-05-0264-A
If an LC method is used, then before the run is started, make sure that there is
enough solvent in the reservoirs for the primary flow rate for all of the sample runs
and the maximum idle time.
5. Select
Leave Mass Spec on in Standby
to keep the mass spectrometer running
after analysis has been completed. This feature allows the heaters and gases to
continue running, even after devices have entered idle state, so that the ion source
and entrance to the mass spectrometer are kept free of contaminants.
6. Select
Fail Whole Batch in Case of Missing Vial
to fail the entire batch when a
missing vial is encountered. If this option is not selected, only the current sample will
fail and the queue will continue to the next sample.
Add Sets and Samples to a Batch
A set can consist of a single sample or multiple samples.
1. In the navigation bar, under
Acquire
, double-click
Build Acquisition Batch
.
2. In the
Sample
tab, in the
Set
list, type a name.
3. Click
Add Set
.
4. Click
Add Samples
to add samples to the new set.
5. In the
Sample name
section, in the
Prefix
field, type a name for the samples in this
set.
Note:
For more information about adding quantitation information to a batch, refer to
the
Advanced User Guide
.
Figure 9-2 Batch Editor Dialog
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...