Operating Instructions — Analyze and Process Quantitative Data
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
105 of 138
Manually Integrate Peaks
Manually integrating peaks should be done last, to limit person-to-person variability. Manually
integrate peaks only if all the peaks have not been found after the algorithm parameters have
been adjusted and updated.
1. In the
Peak Review
pane, click
Manual Integration Mode
.
Figure 11-7 Update Method
Note:
Peaks that are manually integrated, or where the algorithm was changed for
only that peak, are identified as such in the
Record Modified
column of the Results
Table, as are peaks that have algorithm parameter changes for a sample that are not
applied to the entire analyte group.
Figure 11-8 Peak Review Pane: Manual Integration
1
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...