Service and Maintenance Information
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
125 of 138
2. Dampen the lint-free wipe with water and then wipe the front of the orifice plate.
Caution: Potential System Damage: Do not insert a wire or metal brush into
the aperture, to avoid damaging the aperture.
3. Repeat
using the cleaning solution.
4. Wait until the orifice plate is dry.
5. Inspect the orifice plate for solvent stains or lint, removing any residue with a clean,
slightly damp lint-free wipe.
Put the System Back into Service
1. Install the curtain plate on the front end of the mass spectrometer.
2. Install the ion source on the mass spectrometer. Refer to the ion source
Operator
Guide
.
3. Activate the hardware profile.
Empty the Source Exhaust Drain Bottle
Empty the source exhaust drain bottle before it becomes full.
1. Remove the ion source. Refer to the ion source
Operator Guide
.
2. Loosen the clamps connecting the hoses to the cap of the source exhaust drain
bottle.
3. Remove the drain bottle from the cap.
Note:
Persistent spotting or filming is an indicator of contaminated solvent.
WARNING! Radiation Hazard, Biohazard, or Toxic Chemical Hazard:
Deposit hazardous material in appropriately labelled containers. Potential
risk of personal injury if proper procedures for handling and disposing of
hazardous materials are not followed.
WARNING! Hot Surface Hazard: Surfaces of the ion source become hot during
operation. Let the ion source and vacuum interface components cool for at least
30 minutes before starting any cleaning procedures.
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...