Operating Instructions — Hardware Profiles and Projects
4500 Series of Instruments
System User Guide
44 of 138
RUO-IDV-05-0264-A
•
If the device uses
Serial Port
as a communication interface, in the
COM Port
Number
list, select the COM port to which the device is connected.
•
If the device uses
Ethernet
as a communication interface, type the
IP
Address
assigned to the device by the administrator or use the corresponding
Host Name
for the address.
•
If the device uses
GPIB Board
as a communication interface, do not change
the settings for the GPIB board.
The rest of the preset values for the device are probably appropriate. Do not change
them. For information about the
Configuration
and
Communication
tabs, refer to
the Help.
10. To restore the device preset values, on the
Communication
tab, click
Set Defaults
.
11. To save the configuration, click
OK
.
12. Repeat
for each device.
13. In the
Create New Hardware Profile
dialog, click
OK
.
14. To activate the hardware profile, in the
Hardware Configuration Editor
, click the
hardware profile.
15. Click
Activate Profile
.
The check mark turns green. If a red x is shown, then there is an issue with the
hardware profile activation. For more information, refer to
.
16. Click
Close
.
Troubleshoot Hardware Profile Activation
If a hardware profile fails to become active, a dialog appears indicating which device in the profile
failed. A failed profile may be due to communications errors.
1. Read the error message generated. Depending on the message, there may be an
issue with a device or how the communication is set up.
2. Verify that the device has power and is turned on.
3. Verify that the COM port assigned to the device is correct.
Note:
For devices using serial communication, make sure that the serial
port selected matches the serial port to which the device is physically
connected. If a serial expansion cable is used, then the number selected in
the profile is the number on the cable plus two.
Note:
The
Alias
field may also be referred to as the Name box and may be
found on another tab under
Alias
.
Tip!
A hardware profile does not have to be deactivated before activating
another hardware profile. Click a hardware profile and then click
Activate
Profile
. The other profile is deactivated automatically.
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...