Operating Instructions — Analyze and Process Quantitative Data
4500 Series of Instruments
System User Guide
110 of 138
RUO-IDV-05-0264-A
Sample Statistics
Use the
Statistics
window to view the statistics samples, typically for standards and QCs (quality
controls). The data from each available batch in the Results Table opens in tabular form in the
grid and a row of data is shown for each standard or QC concentration.
View the Statistics for Standards and QCs
When viewing more than one Results Table, statistical information about the standards and QCs
for additional batches in the Statistics window can be obtained. This allows the user to compare
results between batches and look for trends in the standards or QCs.
1. With a Results Table open, click
Tools > Statistics
.
2. In the
Statistics Metric
list, select
Concentration
.
3. In the
Analyte Name
field, select an analyte.
4. In the
Sample Type
field, select
Standard
.
The results are shown.
5. Look at the
%CV
and
Accuracy
columns.
The %CV shows the coefficient of variance between the measurements of a single
parameter, for example the area. Accuracy shows how close the plotted point is to
the interpolated value.
Menu
Function
Exclude (Include)
Right-click the point and then click
Exclude
to exclude the point
from the curve. Right-click the point and then click
Include
to
include the dropped point.
Exclude All Analytes
(Include All Analytes)
Right-click a point and then click
Exclude All Analytes
to exclude
all the analytes from the curve. Right-click a point and then click
Include All Analytes
to include the points.
Show Peak
Reviews an individual peak.
Overlay
Overlays two graphs.
Active Plot
Determines which plot is active.
Legend
Shows the graph legend.
Log Scale X Axis*
Uses a log scale for the x-axis.
Log Scale Y Axis*
Uses a log scale for the y-axis.
Delete Pane
Deletes the active pane.
Home Graph
Scales the graph to its original size
* A log scale arranges the data points in a more manageable view so that the effect of all points
can be monitored simultaneously. For this view, select Log Scale Y Axis versus Log Scale X
and not just the log of one axis.
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...