Operating Instructions — Analyze and Process Quantitative Data
4500 Series of Instruments
System User Guide
106 of 138
RUO-IDV-05-0264-A
2. Zoom in on the lower 10% of the peak.
3. Position the cross-hair where the start of the peak is to be defined and then drag the
cross-hair to where the end of the peak is to be defined.
The software shades the area bounded by the base and sides of the peak. Peak
parameters are gray as they are no longer applicable because the peak was drawn
manually.
4. Do one of the following:
•
To make this change permanent, click
Accept
.
•
To discard the changes, clear the
Manual Integration
check box.
Item Description
1
Manual Integration mode
Figure 11-9 Peak Review Pane: Zooming in on a Peak
Item Description
1
Lower 10% of the peak
Tip!
If a peak was correct as originally selected, right-click the peak and
then click
Revert to Method
.
1
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...