July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor
Page 123
5
Binary: xxxxxxxxxx1xxxxx
Hexadecimal: xx2x
Beam2 cell out of range (too far away)
2)
6
Binary: xxxxxxxxx1xxxxxx
Hexadecimal: xx4x
Beam3 cell out of range (too far away)
2)
7
Binary: xxxxxxx1xxxxxxx
Hexadecimal: xx8x
Beam4 cell out of range (too far away)
2)
8
Binary: xxxxxxx1xxxxxxxx
Hexadecimal: x1xx
Beam 1 cell weak signal strength, below weak signal limit 1,
default value -40dB
9
Binary: xxxxxx1xxxxxxxxx
Hexadecimal: x2xx
Beam 1 cell weak signal strength, below weak signal limit 2,
default value -44dB
8 and 9 Binary: xxxxxx11xxxxxxxx
Hexadecimal: x3xx
Beam 1 cell weak signal strength, below weak signal limit 3,
default value -48dB
10
Binary: xxxxx1xxxxxxxxxx
Hexadecimal: x4xx
Beam 2 cell weak signal strength, below weak signal limit 1,
default value -40dB
11
Binary: xxxx1xxxxxxxxxxx
Hexadecimal: x8xx
Beam 2 cell weak signal strength, below weak signal limit 2,
default value -44dB
10 and
11
Binary: xxxx11xxxxxxxxxx
Hexadecimal: xCxx
Beam 2 cell weak signal strength, below weak signal limit 3,
default value -48dB
12
Binary: xxx1xxxxxxxxxxxx
Hexadecimal: 1xxx
Beam 3 cell weak signal strength, below weak signal limit 1,
default value -40dB
13
Binary: xx1xxxxxxxxxxxxx
Hexadecimal: 2xxx
Beam 3 cell weak signal strength, below weak signal limit 2,
default value -44dB
12 and
13
Binary: xx11xxxxxxxxxxxx
Hexadecimal: 3xxx
Beam 3 cell weak signal strength, below weak signal limit 3,
default value -48dB
14
Binary: x1xxxxxxxxxxxxxx
Hexadecimal: 4xxx
Beam 4 cell weak signal strength, below weak signal limit 1,
default value -40dB
15
Binary: 1xxxxxxxxxxxxxxx
Hexadecimal: 8xxx
Beam 4 cell weak signal strength, below weak signal limit 2,
default value -44dB
14 and
15
Binary: 11xxxxxxxxxxxxxx
Hexadecimal: Cxxx
Beam 4 cell weak signal strength, below weak signal limit 3,
default value -48dB
1)
The cell can have been configured to be behind the instrument or very close to the instrument. This can also change
during the deployment dependent on water level changes and sensor tilt.
2)
Cells can have been configured to be too far away. Tilting of the instrument can also result in out of range on some of
the beam cells.