10
11
1
OPTIONAL 3M
™
ATTEST
™
AUTO
‑
READER 490 WEB APP
Web Browser Requirements
The Web Interface is designed to be viewed using an HTML5 compliant browser with
JavaScript enabled (e.g., Google Chrome).
Ethernet Connection:
Connection switch/router with auto
‑
negotiation capability to identify the Auto
‑
reader as
operating at 10 Mbps and half
‑
duplex, or able to be manually configured as 10 Mbps and
half
‑
duplex.
Connecting to the 3M
™
Attest
™
Auto
‑
reader 490 Web App
To use the 3M™ Attest™ Auto
‑
reader 490 Web App, the 3M™ Attest™ 490 Auto
‑
reader
must be connected to the facility network using the supplied Ethernet cable. Consult
with your facility IT personnel for assistance with network connectivity. Each 3M™
Attest™ Auto
‑
reader 490 has a unique MAC Address which is presented on the unit
’
s
LCD Display when the user depresses the Info button three times. The MAC Address is
also provided at the bottom of the 3M™ Attest™ Auto
‑
reader 490 Web App Configuration
page. Communication occurs using the http protocol over Port 80. Users may connect to
the 3M™ Attest™ Auto
‑
reader 490 Web App using a URL based upon either the 3M™
Attest™ Auto
‑
reader 490 serial number or the IP Address. To view the 3M™ Attest™
Auto
‑
reader 490 Web App from a computer on a sub
‑
net different from that to which the
3M™ Attest™ Auto
‑
reader 490 is connected, use the IP Address URL.
Note:
To access the 3M™ Attest™ Auto
‑
reader 490 Web App within a secure network,
the PC must have access to the secure network; options include connecting via a network
Ethernet cable, signing in to the facility
’
s wireless network, or connecting via a virtual
private network (VPN).
A. Serial number URL
1. Connect the supplied Ethernet cable from the 3M™ Attest™ Auto
‑
reader 490 to an
Ethernet jack.
2. Depress the
button once to display the 3M™ Attest™ Auto
‑
reader 490 6
‑
digit
serial number on the left side of the top row of text on the LCD display.
3. Enter http://attestxxxxxx where xxxxxx represents the unit
’
s serial number (e.g. http://
attest101127) from Step 2 in PC web browser address bar and press Enter.
4. 3M™ Attest™ Auto
‑
reader 490 Web App Status page will display on PC. For future
ease of use, add this page to your web browser
“
Favorites
”
.
B. IP Address URL
The 3M™ Attest™ Auto
‑
reader 490 Web App may also be accessed using either a
dynamic or static IP address.
1. Dynamic IP Address
A dynamic IP address is assigned when the Ethernet cable is connected from the 3M™
Attest™ Auto
‑
reader 490 to a live Ethernet jack.
Press and release the
button twice to display the 3M™ Attest™ Auto
‑
reader 490
’
s
dynamic IP address. Enter http://IPaddress (e.g. http://xxx.xxx.xxx.xxx) in the PC
’
s web
browser address bar and press Enter. The 3M™ Attest™ Auto
‑
reader 490 Web App Status
page will display.
2. Static IP address
Some facility IT personnel may require devices connecting to the network to use a Static
IP address. If this is the case at your facility, contact the IT department to request a Static
IP address.
Record the static IP address provided by your facility IT department here: ___________.
You may wish to label the Ethernet jack with the assigned address. Enter http://
StaticIPaddress (e.g. http://xxx.xxx.xxx.xxx) in the PC
’
s web browser address bar and
press Enter. The 3M™ Attest™ Auto
‑
reader 490 Web App Status page will display.
About the 3M
™
Attest
™
Auto
‑
reader 490 Web App Pages
The 3M™ Attest™ Auto
‑
reader 490 Web App includes four pages:
a) Status page
Status
History
3M™ Attest™ Auto-reader
Copyright © 2017 3M, Inc.
ATTEST
ATTEST
1
2
3
4
5
+
–
24
6
7
8
9
10
+
–
SN
100157
2017-09-16 08:31
1. User
‑
entered nickname (optional)
2. Well Configuration
3. Well Number
4. BI Result/Remaining Incubation Time
5. Date and Time
6. 3M™ Attest™ Auto
‑
reader 490 Serial Number
The Status page provides a real
‑
time representation of the 3M™ Attest™ Auto
‑
reader
490 configuration and LCD panel display. The current date and time and 3M™ Attest™
Auto
‑
reader 490 serial number are also indicated on this page. From this page, the user
may navigate to:
• the History page by clicking on History on the left of the screen.
• the Results Detail page by clicking on the well number of an occupied well (i.e. a well
that currently has an incubating BI). Clicking on an unoccupied well takes the user to
the History page.
b) Results Detail page
Copyright © 2017 3M, Inc.
3M™ Attest™ Auto-reader Results Detail
Status
History
Result
Start Time
Well#
Serial Number
BI Type
Control?
BI Lot#
Sterilizer#
Load#
Implant?
Technician
Sterilizer Model
Cycle Type/Cycle
Parameters
Chemical
Integrator?/Indicator?
Notes
−
2012-09-16 07:04:41
#2
100157
This page provides an individual record for an incubated BI. For complete documentation,
the user may print a hard copy and/or save the record as an electronic file.
Note 1:
The top four rows (highlighted in green) are populated automatically by the
3M™ Attest™ Auto
‑
reader 490 and cannot be changed by the user. The remaining rows
are completed by the user to satisfy the documentation requirements of their facility.
Note 2:
The Results Detail page refreshes less frequently that the Status page. Until the BI
result is available, the Results Detail page displays remaining minutes of incubation in the
results field.