NIBP KINKED HOSE
The cuff inflation was too
rapid. There may be kinked
tubing or another air
obstruction.
Check the external NIBP
hose for kinks. If problem
persists, check the routing
of the NIBP tubing inside
the device.
NIBP LOW VOLTAGE
Insufficient voltage is
supplied to the NIBP
module.
Check the module voltage
in the NIBP Test screen.
Inspect the unit or water
ingress. If evidence of
water ingress is found,
allow the water to
evaporate while periodically
checking the module
voltage. If problem persists,
replace the NIBP module.
NIBP MAX PRESSURE
EXCEEDED
Pressure limit exceeded.
If problem persists, do the
following:
•
Check the calibration via
the “NIBP Calibration
Check” on page 37.
•
Check the routing of the
NIBP tubing inside the
device for kinks or
obstructions.
If the NIBP module is out of
calibration, replace the
NIBP module.
NIBP MONITORING
FAILURE
One of the following issues
occurred:
•
A failure was reported by
the NIBP module.
•
The two transducers are
not aligned. They are
either not calibrated,
there is a transducer
failure, or there is a
kinked tube.
•
A calibration problem was
reported by the NIBP
module.
If problem persists, replace
the NIBP module.
NIBP NEONATAL
DETECTED
Device thinks a neonate cuff
is connected. If no neonate
cuff is connected, tubing
may be damaged or kinked.
Verify that neonate cuff is
not connected. If problem
persists, check the routing
of the NIBP tubing inside
the device.
Error Message
Explanation
Technical Action
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Service Manual 9650 0807 01 Rev C...
Page 6: ...TABLE OF CONTENTS iv www zoll com 9650 0807 01 Rev C...
Page 64: ...58 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 102: ...96 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 108: ...CHAPTER 4 REPLACEMENT PARTS 102 9650 0807 01 Rev C Pictures and Diagrams Picture 1 Picture 2...
Page 109: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 103 Picture 3 Picture 4...
Page 110: ...CHAPTER 4 REPLACEMENT PARTS 104 9650 0807 01 Rev C Picture 5 Picture 6...
Page 111: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 105 Picture 7 Picture 8...
Page 112: ...CHAPTER 4 REPLACEMENT PARTS 106 9650 0807 01 Rev C Picture 9 Picture 10...
Page 113: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 107 Picture 11 Picture 12...
Page 114: ...CHAPTER 4 REPLACEMENT PARTS 108 9650 0807 01 Rev C Picture 13 Picture 14...
Page 115: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 109 Picture 15 Picture 16...
Page 116: ...CHAPTER 4 REPLACEMENT PARTS 110 9650 0807 01 Rev C Picture 17 Picture 18...
Page 117: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 111 Picture 19 Picture 20...
Page 118: ...CHAPTER 4 REPLACEMENT PARTS 112 9650 0807 01 Rev C Picture 21 Picture 22...
Page 119: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 113 Picture 23 Picture 24...
Page 120: ...CHAPTER 4 REPLACEMENT PARTS 114 9650 0807 01 Rev C Picture 25 Picture 26...
Page 121: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 115 Picture 27 Picture 28...
Page 122: ...CHAPTER 4 REPLACEMENT PARTS 116 9650 0807 01 Rev C Picture 29 Picture 30...
Page 123: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 117 Picture 31 Picture 32...
Page 124: ...9650 0807 01 Rev C Propaq MD Service Manual 118...
Page 134: ...128 9650 0807 01 Rev C This page intentionally left blank...
Page 138: ...132 9650 0807 01 Rev C This page intentionally left blank...
Page 142: ...This page intentionally left blank...