8.0 Shock Test
WARNING!
Take the necessary precautions to guard against shock or injury before you start
conducting the defibrillator tests.
Keep hands and all other objects clear of the multi-function cable connections and
defibrillator analyzer when discharging the defibrillator.
Before you discharge the defibrillator, warn everyone near the equipment to STAND
CLEAR.
Equipment
Fluke Impulse 7000 DP Analyzer (or equivalent), stop watch
Do This...
Observe This...
Pass/Fail
8.1
Turn on the Analyzer.
8.2
On the Analyzer, select the DEFIB function.
8.3
On the Analyzer, select ENERGY.
8.4 Connect
the
Propaq
MFC connector to the
pads connection.
8.5
On the Propaq unit, apply power to the unit
using a battery or an AC adapter.
8.6
Connect the MFC cable to the Propaq unit.
8.7
Turn on the Propaq unit, and press the Leads
quick access key
(
)
to select
PADS
.
8.8
On the Propaq unit, set energy level to 5J by
pressing the Energy Select DOWN arrow.
8.9
On the Propaq unit, press the
CHARGE
button.
The unit charges to 5J, the red
Shock button illuminates and a
ready tone sounds.
o
o
8.10 On the Propaq unit, press the Shock button.
8.11 On the Analyzer, observe and record the
energy measurement.
Energy is delivered at 5J. Verify
that the delivered energy is within
15% based on the impedance
values of your analyzer in the
table “Delivered Energy at 200J
Defibrillator Setting into a Range
of Loads” on page 24.
o
o
8.12 On the Propaq unit, set the energy level to
50J by pressing the Energy Select UP arrow.
8.13 On the Propaq unit, press the
CHARGE
button.
The unit charges to 50J, the red
Shock
button illuminates and a
ready tone sounds.
o
o
I, II,
III...
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Service Manual 9650 0807 01 Rev C...
Page 6: ...TABLE OF CONTENTS iv www zoll com 9650 0807 01 Rev C...
Page 64: ...58 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 102: ...96 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 108: ...CHAPTER 4 REPLACEMENT PARTS 102 9650 0807 01 Rev C Pictures and Diagrams Picture 1 Picture 2...
Page 109: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 103 Picture 3 Picture 4...
Page 110: ...CHAPTER 4 REPLACEMENT PARTS 104 9650 0807 01 Rev C Picture 5 Picture 6...
Page 111: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 105 Picture 7 Picture 8...
Page 112: ...CHAPTER 4 REPLACEMENT PARTS 106 9650 0807 01 Rev C Picture 9 Picture 10...
Page 113: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 107 Picture 11 Picture 12...
Page 114: ...CHAPTER 4 REPLACEMENT PARTS 108 9650 0807 01 Rev C Picture 13 Picture 14...
Page 115: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 109 Picture 15 Picture 16...
Page 116: ...CHAPTER 4 REPLACEMENT PARTS 110 9650 0807 01 Rev C Picture 17 Picture 18...
Page 117: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 111 Picture 19 Picture 20...
Page 118: ...CHAPTER 4 REPLACEMENT PARTS 112 9650 0807 01 Rev C Picture 21 Picture 22...
Page 119: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 113 Picture 23 Picture 24...
Page 120: ...CHAPTER 4 REPLACEMENT PARTS 114 9650 0807 01 Rev C Picture 25 Picture 26...
Page 121: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 115 Picture 27 Picture 28...
Page 122: ...CHAPTER 4 REPLACEMENT PARTS 116 9650 0807 01 Rev C Picture 29 Picture 30...
Page 123: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 117 Picture 31 Picture 32...
Page 124: ...9650 0807 01 Rev C Propaq MD Service Manual 118...
Page 134: ...128 9650 0807 01 Rev C This page intentionally left blank...
Page 138: ...132 9650 0807 01 Rev C This page intentionally left blank...
Page 142: ...This page intentionally left blank...