20.0 NIBP Calibration Check
Equipment
NIBP Simulator, dual lumen hose w/ NIBP test fixture cuff (9100-3072-TF) or equivalent
Note:
If ECG simulator is connected to the device, either disconnect ECG simulator or
disable the Smart Cuff feature in the Supervisor Menu.
Setup
Attached the hose connected to the T-fitting of the NIBP test fixture cuff to the pressure port of
the NIBP simulator. Set simulator to Static Pressure.
Do This...
Observe This...
Pass/Fail
20.1 Turn on the Propaq unit.
Note
: If the unit is in AED mode, press
the
Manual
quick access key to go into
manual mode.
20.2 Connect a dual lumen hose from the NIBP
simulator to the NIBP port of the Propaq.
20.3 On the Propaq unit, do the following to
enter the NIBP Test control panel:
•
Press the More
(
) then the Setup
(
) quick access keys.
•
Use the navigation keys to select
Supervisor
>
Service
>
Device
Tests
>
NIBP Test
.
Note
: You need a supervisor passcode to
enter the Supervisor menu.
20.4 From the Propaq NIBP Test control panel,
use the navigation keys to select:
•
Close Valve
•
Turn Pump On
20.5 Monitor the pressure reported by the
external pressure standard.
20.6 Once the pressure reaches the target
pressure of 50 +/- 5 mmHg, use the
navigation keys to select
Turn Pump Off
.
20.7 Allow the system to stabilize for at least
10 seconds.
20.8 Verify that the primary transducer agrees
with the reading on the NIBP simulator.
Reading should be within
3mmHg.
o
o
20.9 Verify that the safety transducer agrees
with the reading on the NIBP simulator.
Reading should be within
4mmHg.
o
o
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Service Manual 9650 0807 01 Rev C...
Page 6: ...TABLE OF CONTENTS iv www zoll com 9650 0807 01 Rev C...
Page 64: ...58 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 102: ...96 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 108: ...CHAPTER 4 REPLACEMENT PARTS 102 9650 0807 01 Rev C Pictures and Diagrams Picture 1 Picture 2...
Page 109: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 103 Picture 3 Picture 4...
Page 110: ...CHAPTER 4 REPLACEMENT PARTS 104 9650 0807 01 Rev C Picture 5 Picture 6...
Page 111: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 105 Picture 7 Picture 8...
Page 112: ...CHAPTER 4 REPLACEMENT PARTS 106 9650 0807 01 Rev C Picture 9 Picture 10...
Page 113: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 107 Picture 11 Picture 12...
Page 114: ...CHAPTER 4 REPLACEMENT PARTS 108 9650 0807 01 Rev C Picture 13 Picture 14...
Page 115: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 109 Picture 15 Picture 16...
Page 116: ...CHAPTER 4 REPLACEMENT PARTS 110 9650 0807 01 Rev C Picture 17 Picture 18...
Page 117: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 111 Picture 19 Picture 20...
Page 118: ...CHAPTER 4 REPLACEMENT PARTS 112 9650 0807 01 Rev C Picture 21 Picture 22...
Page 119: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 113 Picture 23 Picture 24...
Page 120: ...CHAPTER 4 REPLACEMENT PARTS 114 9650 0807 01 Rev C Picture 25 Picture 26...
Page 121: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 115 Picture 27 Picture 28...
Page 122: ...CHAPTER 4 REPLACEMENT PARTS 116 9650 0807 01 Rev C Picture 29 Picture 30...
Page 123: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 117 Picture 31 Picture 32...
Page 124: ...9650 0807 01 Rev C Propaq MD Service Manual 118...
Page 134: ...128 9650 0807 01 Rev C This page intentionally left blank...
Page 138: ...132 9650 0807 01 Rev C This page intentionally left blank...
Page 142: ...This page intentionally left blank...