C
HAPTER
2
T
ROUBLESHOOTING
56
9650-0807-01 Rev. C
SERVER
AUTHENTICATION FAILED
The reports server rejected
the conncetion request from
the Propaq MD device.
To correct the problem,
verify the following:
•
The customer ID and
password are correctly
entered in the reports
configuration. (The
password is case
sensitive.)
•
The reports
configuration has the
correct server hostname
and port.
•
The communications
processor (CP) software
is up to date.
SERVER UNREACHABLE
Although a WAN connection
was established, the reports
server connection could not
be established or the
connection was lost during a
transfer.
Retry the operation.
SERVICE ERROR
<CODE>
While communicating with
the remote server, there was
an error.
Retry the operation.
SNAPSHOT SHORT
Snapshot key is stuck.
Replace the front panel.
SPO2 COM ERROR CODE
Communication error.
Check communication
cables. If problem persists,
replace the SpO
2
module.
SPO2 DEVICE FAILURE
Communication failure or
device malfunction.
Check the SpO
2
communication cables. If
problem persists, replace
the SpO
2
module.
SPO2 SENSOR FAILURE
Defective sensor or sensor
error.
Replace the SpO
2
sensor.
SPO2 UNKNOWN
RECOVERY NUM
Recovery error.
Replace the SpO
2
sensor.
SW SHUTDOWN
WARNING
Indication that the battery
has reached the software
shutdown capacity.
Charge the battery.
TEMP1 (or TEMP2)
CIRCUIT FAILURE
The software detected an
out-of-range A/D converter.
Remove probe then power
cycle. If problem persists,
replace the main board.
Error Message
Explanation
Technical Action
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Service Manual 9650 0807 01 Rev C...
Page 6: ...TABLE OF CONTENTS iv www zoll com 9650 0807 01 Rev C...
Page 64: ...58 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 102: ...96 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 108: ...CHAPTER 4 REPLACEMENT PARTS 102 9650 0807 01 Rev C Pictures and Diagrams Picture 1 Picture 2...
Page 109: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 103 Picture 3 Picture 4...
Page 110: ...CHAPTER 4 REPLACEMENT PARTS 104 9650 0807 01 Rev C Picture 5 Picture 6...
Page 111: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 105 Picture 7 Picture 8...
Page 112: ...CHAPTER 4 REPLACEMENT PARTS 106 9650 0807 01 Rev C Picture 9 Picture 10...
Page 113: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 107 Picture 11 Picture 12...
Page 114: ...CHAPTER 4 REPLACEMENT PARTS 108 9650 0807 01 Rev C Picture 13 Picture 14...
Page 115: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 109 Picture 15 Picture 16...
Page 116: ...CHAPTER 4 REPLACEMENT PARTS 110 9650 0807 01 Rev C Picture 17 Picture 18...
Page 117: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 111 Picture 19 Picture 20...
Page 118: ...CHAPTER 4 REPLACEMENT PARTS 112 9650 0807 01 Rev C Picture 21 Picture 22...
Page 119: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 113 Picture 23 Picture 24...
Page 120: ...CHAPTER 4 REPLACEMENT PARTS 114 9650 0807 01 Rev C Picture 25 Picture 26...
Page 121: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 115 Picture 27 Picture 28...
Page 122: ...CHAPTER 4 REPLACEMENT PARTS 116 9650 0807 01 Rev C Picture 29 Picture 30...
Page 123: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 117 Picture 31 Picture 32...
Page 124: ...9650 0807 01 Rev C Propaq MD Service Manual 118...
Page 134: ...128 9650 0807 01 Rev C This page intentionally left blank...
Page 138: ...132 9650 0807 01 Rev C This page intentionally left blank...
Page 142: ...This page intentionally left blank...