CareCube™ Wound Treatment System  

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Inflatable Cushion for Lifting and Supporting  

of Legs during Wound Treatment 

 

 
 

 

 

 

 

Summary of Contents for CareCube

Page 1: ... Aps Hejreskovvej 18 B C DK 3490 Kvistgård Denmark Tlf 45 4913 8855 www winncare dk Aug 2015 User s Manual CareCube Wound Treatment System Inflatable Cushion for Lifting and Supporting of Legs during Wound Treatment ...

Page 2: ...f the CareCube System 9 11 Annual Service Inspection of the CareCube System 9 12 Components in the CareCube Wound Care System 10 13 The Compressor Inserting and removing the air hose 11 14 The CareCube Cushion Manual deflation 12 15 The CareCube Cushion Inserting and removing the air hose 12 16 Inflating the CareCube Cushion 13 17 Deflating the CareCube Cushion with the compressor 14 18 Trouble sh...

Page 3: ...h air from a small compressor which can both inflate and deflate the CareCube The deflated CareCube is only a few centimetres thick making it easy to slide underneath the patient s foot or leg When activated the compressor fills the CareCube completely in approximately 20 seconds lifting the patient s leg to the desired height During treatment the patient s leg rests safely and comfortably on the ...

Page 4: ... be followed meticulously All staff should be trained prior to using the product We recommend that all operators practice using the CareCube and that they try having their legs lifted themselves so as to know what the patients will experience Before each treatment procedure the nurse should think the entire process through and explain it to the patient Symbols used 3 Disclaimer Read this manual ca...

Page 5: ...irective 93 42 EEC invalid 4 Safety Before using the CareCube the caregiver must inspect the cushion and the compressor and make certain that they are fully functional The product must be taken out of service immediately if faults which could have an impact on safety are discovered In that case the CareCube System must be serviced by an authorized service technician before being taken into use aga...

Page 6: ...Identification of this unit Storage Information about storage Keep dry Information about storage 25 90 Relative humidity Identification of safe humidity 0 50º Celsius Safe usage temperature 250 1200 hPa Safe air pressure CareCube Wound Care System Serial No CareCube1 2015 02 100 Storage Keep dry 25 90 relative humidity 0 50 Celsius 250 1200 hPa ...

Page 7: ...of the ingression protection level of the compressor and the cushion respectively 7 Technical Data Materials Cushion PVC Polypropylene Compressor PP electronics IP protection level cushion IPX6 IP protection level compressor IPX1 Maximum continuous usage 10 minutes Maximum noise level 80 DB at a distance of 1 metre Length of net cable 3 metres Power supply 230V 50Hz 120V 60Hz Power 400W Maximum ai...

Page 8: ...r thorough cleaning the CareCube cushion may be cleaned in automated washing machines or with saturated steam for up to 1 minute at 90 Celsius 194 Fahrenheit for 5 minutes at 85 Celsius 185 Fahrenheit or 10 minutes at 80 Celsius 176 Fahrenheit This can only be done with the valve cover inserted and closed CareCube compressor The compressor is electrically operated and runs on either 230V or 120V A...

Page 9: ... the CareCube System The CareCube System does not need any annual service inspection All users should be trained to carry out a daily inspection of the CareCube System before using it Check the following points Never clean the compressor with the power cable plugged in This can lead to electrical shock or danger of fire if alcohol is used for cleaning Never clean the CareCube System with substance...

Page 10: ...2 Components in the CareCube Wound Care System The CareCube Wound Care System consists of 4 components 1 The CareCube Cushion 2 The CareCube Compressor with 3 meter power cable 3 An air hose 4 A backpack for transport ...

Page 11: ... hose from one end to the other If the air hose is inserted in the mounting marked with the word OUT and a green dot air is blown out and the CareCube cushion is inflated If the air hose is inserted in the mounting marked with the word IN and a red dot air is sucked in and the CareCube cushion is deflated In order to insert the air hose in either mounting the black adaptor of the air hose is simpl...

Page 12: ...shed again the valve closes again This function may be used to deflate the CareCube cushion manually 15 The CareCube Cushion Inserting and removing the air hose Push the black adaptor into the valve This locks the air hose in place and automatically opens the valve In order to remove the air hose twist the black adaptor on the air hose a bit sideways and pull it out of the valve This automatically...

Page 13: ...ending on which zone should be accessible for treatment If the patient is incapable of lifting the foot or bending the leg the CareCube Cushion be slid in place using a slide sheet or similar Follow these steps in order to lift 1 Chose lifting height and empty the cushion so that the desired color points upwards 2 Place the empty CareCube Cushion under the desired spot of the patient s foot leg 3 ...

Page 14: ... and or the highest patient comfort 17 Deflating the CareCube Cushion with the compressor Follow these steps in order to deflate the CareCube Cushion 1 Insert the power cable into the socket 2 Insert the air hose in the mounting on the compressor marked with the word IN and a red dot 3 Insert the air hose into the valve on the CareCube Cushion 4 Turn on the compressor using the switch 5 The CareCu...

Page 15: ...osing the valve manually to make certain that it works The valve is defective Contact your for service The Cushion loses air quickly Try opening and closing the valve manually to make certain that it works The cushion has been punctured Contact your dealer for service 19 Warranty The manufacturer offers a 2 year warranty on the CareCube Wound Treatment System for failures caused during production ...