11-36
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019
Working With Menus
key, but the system will automatically take the first and subsequent
measurements based on the o’clock as per the system time.
-
The chosen interval is displayed as “NBP Interval – N Minutes” in the
system message area, where N is the value selected.
To stop a particular measurement within NBP Interval Mode, simply press the
NBP START/STOP key. However, “NBP Interval – N Minutes” will continue to
be displayed in the message area, and another measurement will automatically be
initiated at the next interval. To fully disable NBP Interval Mode, you must re-
enter the Setup NBP Interval menu and change the Interval to Off.
The NBP Interval setting is remembered after a Standby or a power cycle with the
same patient. However, the PPM3 does not automatically initiate NBP Interval
Mode measurements after a Standby or a power cycle until the NBP
START/STOP key is pressed. This tells the PPM3 that the cuff has been applied
to the patient and that NBP monitoring should commence. Once the key has been
pressed, the interval will display in the message area and the measurements will
proceed thereafter based on the o’clock as per the system time.