24-6
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019
Technical Data
Cuff
Infant, Child, Small Adult, Adult, Large Adult
Derived Parameters
Systolic, Mean, Diastolic
Resolution
1 mmHg
Measurement Range
Systolic: 30 to 250 mmHg
Mean: 20 to 230 mmHg
Diastolic: 10 to 210 mmHg
Measurement Accuracy
Complies with AAMI SP10
Transducer Accuracy
± 3 mmHg
Pulse Rate Range
30 to 240 bpm
Pulse Rate Accuracy
± 5% or ± 2 bpm, whichever is greater
Update Interval
Upon measurement completion
Measurement Time
30 seconds (typical)
< 135 seconds (maximum)
Initial Cuff Pressure
160 mmHg (user-selectable)
Repeated Cuff Pressure
Previous sy 40 mmHg
Static Cuff Pressure Accuracy
± 3 mmHg
Overpressure Cutoff
290 ± 3 mmHg (normal means), 300 ± 10 mmHg (back-up)
Measurement Modes
Single Measurement or Auto (Interval) Measurement
Auto Measurement Settings
Off, 3, 5, 10, 15, 30, 60, 120 minutes
Oridion CO
2
/ Capnography
OEM Board
Oridion miniMediCO
2
Method
Sidestream (Non-dispersive IR)
Units
mmHg
Parameters
ETCO
2
, FICO
2
, RRc, IPI
CO
2
Measurement Range
ETCO
2
& FICO
2
: 0 to 150 mmHg
Summary of Contents for PPM3
Page 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Page 4: ......
Page 16: ......
Page 28: ......
Page 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Page 52: ......
Page 96: ......
Page 100: ......
Page 138: ......
Page 160: ......
Page 164: ......
Page 184: ......
Page 194: ......
Page 196: ......
Page 198: ......