N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
10-3
Temperature Monitoring
10.2. Temperature Monitoring Messages
Note
– Factory default Temperature alarm and limit settings may be found in the Menu
Settings table in the PPM3 Monitor Settings chapter.
Message
Parameter
Value
Possible Cause
Suggested Action
TEMP < [lower limit]
[number]
The patient's
temperature has fallen
below the current lower
alarm limit.
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
TEMP > [upper limit]
[number]
The patient's
temperature has risen
above the current upper
alarm limit.
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
TEMP unplugged
[blank]
Temperature probe
disconnected
Check to make sure the temperature
probe is connected to the temperature
cable.
Check to make sure the temperature
cable is connected to the PPM3.
TEMP out of range
---
The patient's
temperature has risen
above the maximum
value the monitor can
accurately detect.
There is a problem with
the connections or with
the hardware.
Check the patient and provide any
necessary clinical care.
Check the temperature cable
connections.
Turn the monitor off, then on.
If message persists, contact Zoe
technical support.
TEMP problem detected
---
Monitor has detected a
hardware problem.
Turn the monitor off, then on.
If message persists, contact Zoe
technical support.
Summary of Contents for PPM3
Page 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Page 4: ......
Page 16: ......
Page 28: ......
Page 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Page 52: ......
Page 96: ......
Page 100: ......
Page 138: ......
Page 160: ......
Page 164: ......
Page 184: ......
Page 194: ......
Page 196: ......
Page 198: ......