21-2
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019
Cleaning
Part
Recommended cleaning method
Reusable SpO
2
Sensor
Materials
•
70% isopropyl alcohol pad
Procedure
Remove sensor from patient and disconnect from
sensor cable. Wipe off with alcohol pad. Allow sensor
to dry before placing it on a patient.
Masimo IRMA (Mainstream)
CO
2
Probe
Materials
•
Cloth moistened with 70% (max) ethanol or
isopropyl alcohol
Procedure
Remove probe from patient and disconnect from
airway adapter. Wipe off with cloth. Allow probe to
dry before placing it on a patient.
Masimo ISA (Sidestream) CO
2
Analyzer
Materials
•
Cloth moistened with 70% (max) ethanol or
isopropyl alcohol
Procedure
Remove probe from patient. Keep sampling line
connected while cleaning to prevent liquids and dust
from entering the analyzer via the LEGI connector.
Wipe off with cloth. Allow analyzer to dry before
placing it on a patient.
CAUTION
– Always disconnect the PPM3 from AC mains before cleaning.
CAUTION
– Do not use harsh chemicals for cleaning – in particular, do
not use disinfectants that contain phenol as they can spot plastics. Do not
steam autoclave, gas sterilize, irradiate, subject to intense vacuum, or
immerse in water or cleaning solution. Be careful to avoid getting cleaning
liquids into connectors or the unit. If this occurs, allow the unit to dry in
Summary of Contents for PPM3
Page 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Page 4: ......
Page 16: ......
Page 28: ......
Page 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Page 52: ......
Page 96: ......
Page 100: ......
Page 138: ......
Page 160: ......
Page 164: ......
Page 184: ......
Page 194: ......
Page 196: ......
Page 198: ......