Section 7.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 89 of 220
Time from Hand foot switch activation
7.9.1.1.
CONTINUOUS FLUOROSCOPY
To generate an image in the continuous pulse fluoroscopy mode, do the following:
• Press
the
Fluoroscopy
button.
The preset selected organ/anatomical program is indicated on the touch control panel.
•
Press the desired organ /anatomical program.
• Initiate
radiation.
For each individual organ program, certain noise filter factors can be set for the continuous
fluoroscopy mode
. This is done by selecting the level from the touch screen.
•
Please contact your service engineer if you wish to set or modify the filter factors for the
low, med, high, levels.
7.9.1.2.
PULSE FLUOROSCOPY
In pulse
fluoroscopy mode
, the system emits radiation pulses as long as you press
the radiation hand or footswitch. Rate < 3 F/sec.
The pulse rate is fixed. The lower the pulse rate setting, the lower the radiation dose per second.
Pulsed fluoroscopy reduces the effective dose rate. Press the
Pulsed Fluoroscopy
button on the
mobile stand and then press the hand or foot switch. Radiation is released for an interval of time,
approximately, one to maximum of 3 pulses per second depending on program time set for
pulses.
Press the
Fluoroscopy
button on the mobile stand keyboard or press the
Pulsed Fluoroscopy
button
on the touch panel to return to standard fluoroscopy. When programmed by a service
engineer, pulsed fluoroscopy can make up to three pulses per second.
•
Please contact your service engineer if you wish to set or modify the
pulse rate factor.
Snap Shot
Pulse
Fluoroscopy
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...