Section 4.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 43 of 220
WARNING:
Never operate the C-arm system if inclination of the floor exceeds 5°.
Death or
serious injury
may result if the
ZIEHM QUANTUM
is operated with a tilt of greater
than 5°
4.5
SWITCHING ON THE SYSTEM
WARNING:
Do not operate the system until the equipment has reached a safe operating temperature
of +15°C to +35°C with no condensation present. Otherwise severe equipment damage
may result.
The C-arm stand has two buttons for switching the system ON or OFF See Fig 4.2 for location of
the two switches. Buttons numbered 3 and 4.
Fig 4.2: C-Arm Mobile stand control keyboard
•
Make sure that the inclination of the system does not exceed 5° from the
level in operating position.
•
Ensure that all electrical connections are properly established.
•
Put on suitable protective clothing.
•
Switch on the system using the Power ON buttons on the mobile stand
C-arm. The button switches on or off all the systems components
simultaneously.
4.5.1
CUSTOMER-SPECIFIC INITIAL SETTINGS
The default settings after power-up vary from system to system, according to the customer-
specific setup.
Furthermore, you may choose to have a certain program and various live image settings preset as
default after power-up. You cannot make these individual pre-settings yourself.
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...