Section 5.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 54 of 220
WARNING
Release the mechanical brakes only for positioning.
Make sure there is enough room to maneuver the C-arm stand and Monitor
assembly before positioning!
Exercise extreme caution when releasing mechanical locks or brakes. To prevent
unwanted movement of the mechanical c-arm assembly, always maintain control
by holding on to and then slowly releasing the mechanical locks before moving
any of the mechanical components of the
ZIEHM QUANTUM
mobile C-arm.
Failure to observe this warning may result in serious injury to patient or operator.
5.3.1
ORBITAL ROTATION
The C-arm can be rotated in an orbital motion by 135° for 23 cm I.I.: –90° from vertical to
horizontal position and +45° forward. A scale with 5° divisions on the outside of the C-arm
profile as well as a mechanical detent at the ‘zero’ position are provided to facilitate precise
positioning.
Fig 5.4 Orbital rotation of the C-Arm
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...