Section 8.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 100 of 220
8.9
MONITOR L TO R
Press the
Center of Image
of the touch screen to transfer
image from the left/live monitor to the reference or right
monitor.
The image can only be transferred from left to right if the
image was not yet stored or recalled, (live transit image).
In certain conditions the transfer of the image will be
inactive for a particular mode or function.
8.10
MONITOR R TO L
Press one of the four
Image Icon’s
on right side of the touch screen once to
select image and recall image to the right monitor then touch the image icon
a second time after a 1/2 to 1 sec delay to transfer image from the reference
or right monitor to the live or left monitor.
In certain conditions the transfer function may not be active for particular
modes or functions.
8.11
INV VIDEO
The INV Video function can be used with both stored images and real-time fluoroscopy.
Press the
INV Video
button to reverse the light and dark areas of the image and
create an image that resembles a film negative.
The effect can be changed latter by recalling the image and pressing the
INV
button
again.
When you press store button the image will be stored with the
INV
image attribute
so the image will be recalled with the Black and White levels inverted.
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...