Section 8.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 104 of 220
8.14.1.
KV DISPLAY
The default setting is 75 kV. During fluoroscopy, the
ZIEHM QUANTUM
adjusts kV levels
automatically using dose rate control circuitry, unless manually overridden. After fluoroscopy,
the last kV value used is stored in the system and is shown in the kV section of the display
window. See left display window above for sample of display values.
8.14.2.
MA DISPLAY
The
ZIEHM QUANTUM
controls mA level in relation to kV level. The mA value is displayed
in the left window during live fluoroscopy.
The last value used by the system is stored and recalled when the system returns to fluoroscopic
mode. This value will continue to be used until either new kV values are entered or until a new
operating mode is selected. See left display window above for sample of display values.
8.14.3.
EXPOSURE TIME(S)
In Radiographic mode, the system sets the current to 20 mA and an exposure time from 0.1 to 4.0
seconds can be selected using the
Up/Down Arrow
button. This exposure time is then stored and
displayed. See left display window above for sample of display values.
8.14.4.
CGYCM2 / ERROR
The
cGycm
2
display window has two functions one to show dose area product value in Grays.
The other function provides the user with a way to see system error codes. The dose display
function is not operational on the U.S. models. The error codes however will be displayed in this
window. See right display window above for sample of display values.
8.14.5.
MIN FLUOROSCOPY TIME DISPLAY
Fluoroscopy and radiography time are displayed in minutes and seconds. Every five
minutes of fluoroscopic time an alarm will sound to warn the user that 5 minutes of
fluoroscopic time has elapsed. This will continue every five minutes. The sound will
not stop until the operator reset the alarm.
See right display window above for sample of display values.
8.14.5.1.
AUDIBLE ALARM
After five minutes of radiation have accumulated, an alarm is triggered and an LED in
the
Min
button begins to flash and the
Reset Timer
button turns red on the DeskView.
To turn off the alarm, press the
Min
button on the mobile stand or the
Reset Timer
button on the DeskView touch screen. The LED and the Reset button will continue to
flash on the mobile stand; the touch screen button will turn red. To turn off both the
alarm and the LED, press and hold the
Min
or
Reset Timer
button.
CAUTION
If you do not switch off the alarm after 30-60 seconds max., radiation could be
terminated automatically (Can be set by service technician)
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...