Section 9.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 125 of 220
9.0
CINE & DIGITAL SUBTRACTION MODE
DSA Mode of Subtracted Angiography
The
DSA
functionality.
9.1
DSA/CINE FUNCTIONS
Subtracted Dynamic Image Acquisition (DSA/CINE)
Digital Subtraction Angiography (DSA)
DSA
is always performed by storing a sequence of images captured at a specified rate, between 1
to 15 Frames per second. The
DSA
function is used to acquire high quality dynamic subtracted
images, which is important when viewing Pain Management images of the spine and simple
peripheral injections.
When
DSA
is selected, the imaging system optimizes the CCD camera’s operating parameters for
subtracted image acquisition, and the imaging system selects values for LUT, Window/Level,
Noise Reduction, and other parameters which make the image suitable for DSA.
Frame Rates
of 1 to 15 Frames Per Second (FPS) can be selected, depending on the options
included with the system and on the rate that is appropriate for the medical procedure.
During
DSA
the system will automatically display on the reference monitor a representative
native image with some of the
DSA
special settings for CCD camera, making the image on the
reference monitor a little flatter than a normal fluoroscopic image but provides a live native un
subtracted image during DSA acquisition. The subtracted (DSA) image will be displayed on the
live monitor. This provides the operator with both a live un-subtracted and subtracted image at
the same time.
Optionally:
The user may have the service installer select in the service mode one of two Input
Look Up Tables and set Window and Level values to suit individual needs of image contrast and
brightness. Factory defaults are set to normal viewing range for most operators.
Optionally:
When selected for operation, the
DSA
injection of the vascular anatomy is normally
fully opacified when the operator releases the foot switch or hand switch. The reference monitor
will continue to display the regions of the vascular anatomy that had the greatest contrast during
injection; this is considered a Maximum Opacified image.
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...