13.0
INSPECTION & MAINTENANCE
13.1
ROUTINE CHECKS TO BE PERFORMED BY THE USER
13.1.1
SYSTEM INSPECTION
Every month, the System User is responsible for performing the inspections listed in the table
below.
13.1.2
TABLE OF REGULAR CHECKS
Check
Interval
Remarks
X-ray generator
Monthly
Check for physical damage.
Radiation switches
Monthly
Must
initiate radiation only if pressed
permanently. Releasing them must terminate
radiation.
After 1 sec max. (Depending on the stack filter
settings).
Radiation signals
Monthly
During the exposure, the yellow
Radiation warning lamp on top of the monitor
housing and the
X-ray symbol
on the control
panel must be illuminated.
AKR Air Kerma Display
Monthly
Dose display in rate mGy/min and accumulative
mGy must be within 10 % of pervious monthly
test values
Audible alarm
Monthly
Must sound in direct radiography mode during the
whole exposure time, and in fluoroscopy mode
after 5 min.
Information labels
Monthly
All warning and information labels must be
properly attached and easily legible.
Power cable
Monthly
Must not show any signs of physical damage.
C-arm stand wheels
Whenever necessary
Clean when dirty.
NOTE:
A detailed maintenance schedule can be found in the “
ZIEHM QUANTUM
Technical
Manual”. Technical information necessary to repair or upgrade the
ZIEHM
QUANTUM
system will be made available by
Ziehm Imaging, Inc.
to authorized and
qualified personnel upon request.
13.2
TEST RECORDING
Consistency test according to national regulations for dose level.
At regular intervals, at least once a month, you must check whether the system shows any
deviations from the reference values by performing a consistency test.
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...