Section 7.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 85 of 220
7.0
STANDARD FLUOROSCOPIC OPERATIONS
7.1
GENERAL DESCRIPTION
The
ZIEHM QUANTUM
family of x-ray systems has digital processing systems that are
designed to extend the advantages of digital image technology to C-Arm Fluoroscopy. Their
image enhancement capabilities answer the specific needs of a variety to clinical fluoroscopic
applications.
The
ZIEHM QUANTUM
is a mobile x-ray system developed for surgical and diagnostic
applications such as fluoroscopy and spot-film radiography. It is not to be used in the presence of
flammable anesthetics or other explosive gases.
7.2
SPECIAL FEATURES
The
ZIEHM QUANTUM
has been engineered to provide superior quality and performance.
Here are some of its outstanding features:
•
High-frequency generator that produces a precise spectrum of radiation capable of superior
penetration.
•
Efficient generator design that allows the
ZIEHM QUANTUM
to perform fluoroscopy at
maximum kV for nearly 20 minutes.
•
A 1/2 Dose function that reduces the working dose rate by approximately half.
•
Optimum adjustment for various diagnostic requirements made possible by the system’s
Anatomical programming.
•
Reducing dose for extended procedures through pulsed fluoroscopy. (<3 F/sec).
•
Digital memory featuring digital image rotation, and 16 bit image processing.
•
Noise filtering is available through Frame integration.
•
Greater resolution is achieved through the image intensifier’s electronic focus and
magnification.
•
Easier system positioning and maneuvering, due to the compact design of the image
intensifier and unique monitor mounting system.
•
Easier cleaning, clean design and unique cable arrangement.
•
Quick access to fluoroscopy in less than 1 minute from powering on.
7.3
IMAGE SYSTEM DESIGN
The
ZIEHM QUANTUM
image processing system is a modular “FSRU” (Field Service
Replaceable Unit) and allows for quick repair and upgrade. Components were selected for their
reliability to promote long service life.
7.4
IMAGING PROCESSING
The imaging system is an integral part of the C-Arm and is designed to work hand-in-hand with
the C-Arm control. The image processor is activated by the foot switch or hand switch and is
controlled by the C-Arm control panel or the DeskView touch panel. Video is acquired through
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...