Section 7.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 88 of 220
•
Grayscale Inversion
: To display an image with negative grayscale
•
Iris Collimator:
Touch control for iris collimator allows tight collimation
•
Slot Collimator:
Touch control for closing off the blooming effects at the edge of the
image in hip and long bone procedures
•
Change orientation of image on the live monitor:
-
Reverse Top/Bottom
: Vertical image
-
Reverse Left/Right
: Horizontal image reversal
-
Image Rotation:
Rotate image to align the image for correct viewing
NOTE
Any adjustments that are applied to a live image on the live monitor remain valid
for all subsequent live images until you choose other settings.
NOTE
Any adjustments that are made to a recalled or none live image will only apply
to that image and will not apply when radiation is started again.
NOTE
When you save an image, it will be saved with all rotations, reversals, filter
settings, and markers, etc. All the modifications to the live image when stored
are visible when the image is displayed as a thumbnail in the mosaic.
NOTE
When you save an image, it will be displayed on the reference (right) monitor as
a thumbnail image on the lower part of the screen as well as on the larger display
area.
7.9
FLUOROSCOPY MODES
7.9.1
CONTINUOUS AND PULSE FLUOROSCOPY
Both fluoroscopy modes work with automatic exposure rate control (AERC), unless manual
exposure rate setting has been selected explicitly by the user.
The tube voltage and the tube current are adjusted automatically, taking into account the selected
fluoroscopy program as well as the thickness and structure of the object.
Automatic exposure rate control (AERC) is active on all three Fluoroscopic modes, and reduces
the radiation burden of both patient and operating staff to a minimum and prevents overexposure
of the screened body region.
Summary of Contents for QUANTUM
Page 1: ...ZIEHM QUANTUM USER S MANUAL...
Page 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Page 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Page 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Page 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Page 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Page 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Page 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Page 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Page 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...