![background image](http://html1.mh-extra.com/html/zeiss/s8/s8_instructions-for-use-manual_3317143118.webp)
118
Operation
G-30-1346-en
S8 Floor Stand
Issue 4.0
Printed on 05. 06. 2001
Failure of lamp control
•
Press the Manual key (1) if brightness control is no longer possible.
Note:
When the manual function has been activated, all electrical control sys-
tems are inoperative. The lamp brightness is automatically adjusted to a
fixed setting.
Failure of focusing system
•
Press Manual key (1) if, for example, the focusing system always
moves into its upper or lower end position.
•
Focus by moving the suspension arm of the ceiling mount (floor stand)
up or down.
Failure of magnetic clutches
If the magnetic clutches fail (magnetic clutches are locked), you can man-
ually position the articulated arm including the microscope by overcoming
the locking effect of the magnetic clutches.
Summary of Contents for S8
Page 1: ...S8 Floor Stand Instructions for use G 30 1346 en Issue 4 0 Printed on 05 06 2001...
Page 2: ......
Page 15: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 15...
Page 17: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 17 1 1 2 3...
Page 19: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 19 3 2 1 4...
Page 21: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 21 2 3 1 4...
Page 23: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 23 1...
Page 25: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 25...
Page 26: ...26 Safety G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 28: ...28 Description G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 31: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 31 1 5 3 4 6 2...
Page 33: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 33 1 2...
Page 35: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 35 1 2 3 4 5...
Page 37: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 37 S2 3 4 5 6 7 1 2 8...
Page 39: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 39 1 2 5 3 4 6 6...
Page 43: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 43 8 9 1 2 3 4 5 6...
Page 45: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 45 7 9 8 10...
Page 47: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 47 2 4 3 1...
Page 49: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 49 5 8 6 7...
Page 50: ...50 Description G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 62: ...62 Preparations for use G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 65: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 65 1 2...
Page 67: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 67 1...
Page 71: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 71...
Page 113: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 113...
Page 115: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 115 7 1 2 3 4 5 6...
Page 117: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 117 2 3 1 4...
Page 119: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 119 1...
Page 120: ...120 Operation G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 146: ...146 Index G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 147: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 147...