![Zeiss S8 Instructions For Use Manual Download Page 123](http://html1.mh-extra.com/html/zeiss/s8/s8_instructions-for-use-manual_3317143123.webp)
G-30-1346-en
S8 Floor Stand
Issue 4.0
Printed on 05. 06. 2001
Maintenance / Further information
123
Insufficient surgical field illu-
mination.
Brightness level set too low. Adjust brightness using dis-
play field of the stand or foot
control panel.
-
Halogen lamp not properly
plugged into lamp socket.
Push halogen lamp into
lamp socket.
-
Defective S light guide (illu-
mination not uniform).
Contact service dept.
Light guide probably has to
be changed.
-
Lamp brightness cannot be
adjusted.
Manual function activated.
(Yellow LED above the key
is on).
Deactivate manual function. -
Motorized focus and zoom
functions of the surgical mi-
croscope are inoperative.
Manual function activated.
(Yellow LED above the key
is on).
Deactivate manual function. -
Suspension arm in hori-
zontal position and cannot
be moved up or down.
Suspension arm still locked. Pull out locking knob and
turn by 180°.
-
In combination with three
successive beeps.
– Error message during
software check, after the
stand has been switched
on.
– Error message indicat-
ing an internal system
error.
Manual operation is pos-
sible.
Contact service dept.
-
Stand wobbles.
Floor not level. Stand base
not ideally positioned.
Turn base slightly. The artic-
ulated arm should be at right
angles to the unstable axis.
Problem
Possible cause
Remedy
See
Summary of Contents for S8
Page 1: ...S8 Floor Stand Instructions for use G 30 1346 en Issue 4 0 Printed on 05 06 2001...
Page 2: ......
Page 15: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 15...
Page 17: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 17 1 1 2 3...
Page 19: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 19 3 2 1 4...
Page 21: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 21 2 3 1 4...
Page 23: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 23 1...
Page 25: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Safety 25...
Page 26: ...26 Safety G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 28: ...28 Description G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 31: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 31 1 5 3 4 6 2...
Page 33: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 33 1 2...
Page 35: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 35 1 2 3 4 5...
Page 37: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 37 S2 3 4 5 6 7 1 2 8...
Page 39: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 39 1 2 5 3 4 6 6...
Page 43: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 43 8 9 1 2 3 4 5 6...
Page 45: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 45 7 9 8 10...
Page 47: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 47 2 4 3 1...
Page 49: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Description 49 5 8 6 7...
Page 50: ...50 Description G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 62: ...62 Preparations for use G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 65: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 65 1 2...
Page 67: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 67 1...
Page 71: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 71...
Page 113: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 113...
Page 115: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 115 7 1 2 3 4 5 6...
Page 117: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 117 2 3 1 4...
Page 119: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 Operation 119 1...
Page 120: ...120 Operation G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 146: ...146 Index G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001...
Page 147: ...G 30 1346 en S8 Floor Stand Issue 4 0 Printed on 05 06 2001 147...