![background image](http://html2.mh-extra.com/html/zeiss/contura/contura_operating-instructions-manual_3323956124.webp)
Errors occurring prior to the measuring operation
8-2
61211-1310402
CONTURA
Errors occurring prior to the measuring operation
If the CMM cannot be started, check whether the following errors can
be excluded:
Errors
Cause
Measure
Check
The CMM cannot be
started.
Wrong start-up se-
quence.
–
Follow the specified or-
der.
–
First, switch on the CMM
and then the computer.
Controller
The controller is off.
Switch the control unit on:
–
Press the lateral
push-but-
ton
on the control con-
sole.
The display must be
visible.
Drives
The drives are off.
Switch on the drives:
–
Press the
key
on the con-
trol console.
The LED above the
key must light up.
The
EMERGENCY
STOP button
on the
control console is
locked.
Unlock the button:
–
Slightly turn the
EMER-
GENCY STOP button
.
The button pops out
a few millimeters.
–
If the above mentioned errors can be excluded, put the CMM out of
operation and repeat the start-up procedure.
–
If measuring operation is still not possible, call a ZEISS service engi-
neer or our support team.
➤
Summary of Contents for CONTURA
Page 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Page 62: ...CMM 4 8 61211 1310402 CONTURA...
Page 68: ...Installation 5 6 61211 1310402 CONTURA...
Page 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Page 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Page 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Page 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Page 130: ...Service features 8 8 61211 1310402 CONTURA...
Page 150: ...2Glossary 2 61211 1310402 CONTURA...
Page 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Page 157: ......
Page 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...