Evaluation of the measured results
7-24
61211-1310402
CONTURA
Source of error
Cause
Workpiece condition
–
The workpiece shows signs of deformation and roughness.
–
The workpiece is soiled.
–
The workpiece is covered with oil, for example drops of oil in the
small bores.
–
The workpiece material is too soft.
Magnetic field
–
A malfunction was caused by a magnetic field.
Possible causes of a magnetic field: for example workpiece, clamp-
ing tool, stylus system extensions.
Qualification
–
Different axis clamping set than for qualification.
–
Different measuring force set than for qualification (depending on
the measuring method, with measuring probes)
Probing
–
Probing has not been performed perpendicular to the probing sur-
face.
Summary of Contents for CONTURA
Page 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Page 62: ...CMM 4 8 61211 1310402 CONTURA...
Page 68: ...Installation 5 6 61211 1310402 CONTURA...
Page 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Page 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Page 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Page 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Page 130: ...Service features 8 8 61211 1310402 CONTURA...
Page 150: ...2Glossary 2 61211 1310402 CONTURA...
Page 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Page 157: ......
Page 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...