Preparations for start-up
61211-1310402
CONTURA
6-9
Attaching the holders to the profile rail
The version of the holder used depends on the probing system. Gener-
ally, fastening is the same for all versions. All holders are fastened to the
profile rail using Allen screws. For VAST holders two screws are required,
while other holders only require one.
The holders are generally preassembled. The procedure for moving, re-
moving or adding a holder is described in the following.
NOTE
With RDS probing systems, two washers are required to fasten the
holder to the profile rail.
The minimum distance between two holders must be 150 mm.
Example of how to mount a holder for the VAST adapter plates:
A Cap
B VAST holder; top view
C VAST holder; front view
1 Slot nut
2 Screw; M6 x16
3 Cover of a holder
4 Brush for cleaning the adapter plate
1
Remove the cap
[A]
.
Summary of Contents for CONTURA
Page 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Page 62: ...CMM 4 8 61211 1310402 CONTURA...
Page 68: ...Installation 5 6 61211 1310402 CONTURA...
Page 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Page 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Page 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Page 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Page 130: ...Service features 8 8 61211 1310402 CONTURA...
Page 150: ...2Glossary 2 61211 1310402 CONTURA...
Page 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Page 157: ......
Page 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...