Malfunctions during measuring operations
8-4
61211-1310402
CONTURA
Malfunction
Cause
Action
Check
Travel move-
ment too slow.
The set travel speed is
too low.
Increase the travel speed:
–
Turn the control knob on the
control console clockwise.
The default value set
in the measuring soft-
ware is too low, for
example 50 %.
–
Set the default speed value in
the measuring software to
100 %.
The SLOW function is
active.
–
Disable the SLOW function via
the control console.
Display on the control
console
Stylus system
change
No rack holder has
been assigned to the
stylus system.
–
Assign a holder to the stylus
system.
The position of the
changer rack or
holder has changed.
–
Reassign the holder.
Collision: missing in-
termediate positions.
–
Reassign the holder and set in-
termediate positions.
Two stylus systems
have been assigned
to one holder.
–
Reassign the holder.
Summary of Contents for CONTURA
Page 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Page 62: ...CMM 4 8 61211 1310402 CONTURA...
Page 68: ...Installation 5 6 61211 1310402 CONTURA...
Page 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Page 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Page 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Page 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Page 130: ...Service features 8 8 61211 1310402 CONTURA...
Page 150: ...2Glossary 2 61211 1310402 CONTURA...
Page 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Page 157: ......
Page 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...