![background image](http://html2.mh-extra.com/html/zeiss/contura/contura_operating-instructions-manual_3323956061.webp)
CMM
61211-1310402
CONTURA
4-7
Example of a reference temperature for measurement of 20°C:
Temperature
Change
Status
Action
Per day
18.6 - 20.0°C
1.4 K
OK
20.0 - 21.5 °C
1.5 K
OK
18.3 - 20.0 °C
1.7 K
Not OK
Wait for 48 hours until the newly
set temperature is constant.
Per hour
19.0 - 20.0°C
1 K
OK
20.0 - 20.9 °C
0.9 K
OK
18.8 - 20.0 °C
1.2 K
Not OK
Wait for 48 hours until the newly
set temperature is constant.
Per meter
19.0 - 20.0°C
1 K
OK
20.0 - 20.9 °C
0.9 K
OK
18.8 - 20.0 °C
1.2 K
Not OK
Wait for 48 hours until the newly
set temperature is constant.