Preparation for measuring operation
61211-1310402
CONTURA
7-13
–
The stylus system components have not been damaged.
Fastening the reference sphere holder to the measuring
table
There is a through hole in the base of the reference sphere holder for an
M12 screw.
1 Reference sphere holder
2 Through hole for M12 screw
–
Place the reference sphere holder on the measuring table and screw
it into a threaded hole in the measuring table (max. torque 35 Nm).
Select the position of the reference sphere holder so that the refer-
ence sphere can be probed with all styli.
If it is not possible to qualify certain styli, you may change the orienta-
tion of the reference sphere. Example:
1 Vertical orientation of the reference sphere
Reference measurement
The position of the reference sphere must be determined before the sty-
lus vector can be determined. Use the master stylus for this purpose.
Proceed as follows:
Summary of Contents for CONTURA
Page 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Page 62: ...CMM 4 8 61211 1310402 CONTURA...
Page 68: ...Installation 5 6 61211 1310402 CONTURA...
Page 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Page 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Page 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Page 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Page 130: ...Service features 8 8 61211 1310402 CONTURA...
Page 150: ...2Glossary 2 61211 1310402 CONTURA...
Page 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Page 157: ......
Page 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...