![background image](http://html1.mh-extra.com/html/zeiss/centermax/centermax_operating-instructions-manual_3322995119.webp)
Preparations for a measuring run
7-19
CenterMax Operating Instructions
61211-1020202
• Screw the metal plate on the measuring table (max. torque
40Nm).
Select the metal plate position so that probing of the probe sphere
is possible with all probes.
• Place the calibration standard on the metal plate and turn the
rotary lever 90° to position
I
in order to activate the magnet.
The calibration standard can be placed in any position on the metal
plate.
Carrying out a reference measurement
The position of the calibration standard must be calibrated prior to
the length of the styli. Use the master probe for this purpose. Proceed
as follows:
1
Insert the master probe in the adapter plate receptacle
➤
page 6-16
.
2
Call the menu for reference measurement
➤
Operating instruc-
tions for the measuring software.
3
Select probe (stylus) 1 on the control panel.
• Press the button between the five LEDs until the LED for probe
1 lights up.
4
Probe the sphere of the calibration standard to determine the
diameter of the sphere.
5
Remove the master probe from the adapter plate receptacle
➤
“Removing the probe (manually)” on page 6-17
.
Now, you can calibrate all probes.
Calibration procedure
After the reference measurement has been carried out, the probes
can be calibrated.
Semi-automatic
calibration should be carried out for all probe sys-
tems.
Tensor calibration
should be carried out for the VAST measuring
probe system. This is important since the measuring force causes a
1
3
2
4
5
NOTE
hss
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......