Aligning the probe
It is still possible to align the probe after it has been assembled. This
may be necessary if some points on the workpiece cannot be probed.
The alignment of the probe for the VAST probe system is described in
the following.
1 Screws in the adapter plate for VAST
• Loosen the three screws in the adapter plate so that the probe
configuration can be turned. Do
not
completely unscrew the
screws!
• Insert the probe in the adapter plate receptacle.
• Turn the probe configuration until the required position is reached.
• Hold the probe configuration and remove the probe from the
adapter plate receptacle.
• Tighten the screws with a torque of 2Nm.
Preventing measuring errors
Measuring errors may have several causes. Answer the following
questions before starting the measurement:
1
Questions regarding the prevention of measuring
errors
yes / no
–
Are the probing surfaces of the workpiece clean?
–
Is the workpiece secured against shifting?
–
Are the probe assembly criteria fulfilled?
–
Are the probe and the stylus in perfect condition?
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......