7-14
Measuring operation
61211-1020202
CenterMax Operating Instructions
Preparations for a measuring run
Prerequisites for a perfect measuring run
The following preparations have to be made to ensure maximum
accuracy of the CMM.
•
Switch on the CMM at least 15 minutes before starting the mea-
suring run.
The CMM must be adapted to room temperature prior to the mea-
suring run.
•
Always carry out a reference measurement after the reference
point travel and calibrate each probe used for measurement.
•
Carry our temperature compensation
➤
page 7-22
.
Reference point travel
Reference point
The reference point or origin is located in the upper right-hand corner
of the measuring volume. It must be determined by a reference point
travel
prior
to the measuring run
➤
“Carrying out a reference point
travel” on page 7-16
.
NOTE
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......