Terminating a measuring run
7-33
CenterMax Operating Instructions
61211-1020202
Terminating a measuring run
Effect of system shutdown
If you put the CMM out of operation and switch it off completely, the
following data will be lost.
–
Reference point
–
Information on the coordinate systems.
When putting the machine into operation again, you must perform
another reference point travel.
Shutdown procedure
After the CMM is put out of operation, the probe head remains in its
last position. The space coordinates of its position are lost.
In order to avoid faults when the CMM is put back into operation, the
measuring run should be terminated correctly. For information on the
operator‘s controls, please refer to
➤
page 2-8
.
• Proceed as follows:
CMM
1
Move the probe head to a safe position.
This position must be inside of the measuring volume. Prepare for
the next start-up by positioning the probe head close to the refer-
ence point.
Computer
2
Terminate the measuring software.
3
Power down the operating system.
4
Switch off the computer.
Control cabinet
5
Switch off the drives.
• Press the quickstop button on the control cabinet and unlock it
again.
6
Switch off the control.
• Set the rotary switch to the "OFF" position.
7
Switch off the power supply.
• Turn the main switch counterclockwise to the "0" position.
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......