Checking the protective circuit
6-9
CenterMax Operating Instructions
61211-1020202
• Move the probe (head) in the x axis at constant speed.
• Try to stop the travel motion (v
x
) by pressing one hand against
the quill housing at the marked position (F).
This increases the shearing force of the drives. Once the shearing
force has exceeded a certain value, all drives will be deactivated.
v
x
: Speed of quill in x axis
F: Force exerted against the quill housing by pressing it with one hand.
Important!
You may press only against the quill housing.
•
Do
not
press against the quill
or
against the probe head.
The probe head and quill could otherwise be damaged.
Check by a ZEISS service
engineer
Under normal circumstances, the force of your hand should suffice to
stop probe movement. If the movement can not be stopped in this
way, the protective circuit must be checked immediately by a ZEISS
service engineer.
• Call your service engineer and arrange for him to check the protec-
tive circuit.
Danger!
If the protective circuit is not functioning properly, the CMM must be
taken out of operation until the protective circuit has been checked by
a ZEISS service engineer.
Reactivating the drives
If you can stop the travel motions, you must reactivate the drives
➤
Operational controls.
v
x
F
!
!
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......