Probing the workpiece
7-27
CenterMax Operating Instructions
61211-1020202
If you answered all of the above questions with "yes“, measuring
errors can be excluded to a large extent.
Special features of VAST probe system
Setting the measuring force
The measuring force has to be set for measuring probe systems such
as VAST. The value of the measuring force must be entered in the
measuring software
➤
operating instructions for the measuring soft-
ware.
The measuring force must be set prior to calibrating the probe. Set
the measuring force for each probe individually if required.
Why does the measur-
ing force have to be
set?
During probing, the probe element causes changes of the workpiece
form. The extent of these form changes influences the precision of
the calculated values. In order to avoid falsified measuring data, the
measuring force should be adapted to the characteristics of the work-
piece material.
–
Has the reference point travel been carried out?
–
Has the reference measurement been carried out?
–
Has the calibration been carried out correctly?
–
Has tensor calibration for the VAST probe system
been carried out?
–
Have all probe styli been calibrated?
–
Has the correct probe stylus been selected?
–
Has the same axis clamping as for calibration been
selected?
–
Has the same measuring force as for calibration been
selected? – Only for VAST probe system.
–
Has temperature compensation been carried out?
Questions regarding the prevention of measuring
errors
yes / no
NOTE
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......