7-20
Measuring operation
61211-1020202
CenterMax Operating Instructions
certain bending of the probe. Tensor calibration makes it possible to
correct the bending for subsequent probings.
Tensor calibration must be carried out by
means of vectors
. The probe
can be moved in all axes.
1
Insert the probe into the adapter plate receptacle
➤
page 6-16
.
2
Open the menu for calibration
➤
operating instructions for the
measuring software.
3
Select the clamping status of the axes, i.e. clamped or unclamped.
The same axis clamping status must be set as will be required later
on for workpiece measurement to prevent measuring errors.
Axis clamping must be performed via the control panel.
4
Probe system specific actions:
• The measuring force must be set for the VAST probe system.
5
Select the stylus to be calibrated on the control panel.
6
Call the menu for calibration in the measuring software and probe
the calibration sphere
➤
Operating instructions for the measuring
software.
The calibration sphere diameter is output as the result.
7
Check the result.
Standard deviation
The result is evaluated based on the standard deviation. There is no
maximum standard deviation which applies to all applications. The
tolerable deviation depends on many factors: e.g. resolution and
precision of the CMM, length and rigidity of the probe shaft, qual-
ity of the probe tip, cleanness of the environment. Normally, the
standard deviation should be within a range of a few micrometers
(e.g. 0.002).
If the result is not satisfactory, repeat the calibration or look for
plausible causes. It might be necessary to decide whether or not to
accept measuring inaccuracies.
8
Repeat points 2 to 6 for all other styli of the probe.
1
3
2
4
5
Summary of Contents for CenterMax
Page 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Page 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Page 10: ...61211 1020202 Operating Instructions Contents 8...
Page 14: ...Foreword 61211 1020202 Operating Instructions...
Page 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Page 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Page 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Page 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Page 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Page 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Page 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Page 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Page 162: ...61211 1020202 Operating Instructions Glossary 2...
Page 164: ...61211 1020202 Operating Instructions Appendix 2...
Page 165: ......