STARTUP
Carl Zeiss
Mounting Optional Components
Axio Scope.A1
66
M60-2-0007 e 05/08
3.4.13
Changing the Filter in the Filter
Wheel Transmitted Light
•
Remove the stage carrier (see chapter
599H599H
3.1.11).
•
Unscrew three screws (
600H600H
Fig. 3-53/
3
) of the field
diaphragm socket (
601H601H
Fig. 3-53/
4
) and pull the
socket out of the stand foot.
•
If necessary, pull the filter to be changed
(
602H602H
Fig. 3-53/
1
) out of the corresponding position
of the filter wheel (
603H603H
Fig. 3-53/
2
).
•
Put the new filter into the position.
A filter position is always designed as
firmly installed shutter position.
•
Repeat the procedure for all filter wheel
positions.
•
Put in the field diaphragm socket and screw it in.
•
Mount the stage carrier (see chapter
604H604H
3.1.11).
Fig. 3-53
Changing the filter in the filter
wheel transmitted light