STARTUP
Carl Zeiss
Mounting the Standard Components
Axio Scope.A1
38
M60-2-0007 e 05/08
(4) Removing and mounting the additional
sleeves
Both gear knobs for the stages with the order no.
432035-9030-000, 432035-9040-000 and
432035-9050-000 are equipped with additional
sleeves for an even more sensitive adjustment of
the object position. These sleeves can be removed
in those cases when it is important to shift the
object more quickly.
•
Loosen both clamping screws (
349H349H
Fig. 3-13/
4
) on
the lower additional sleeve (
350H350H
Fig. 3-13/
3
) and
remove them by pulling them down. Then
loosen both clamping screws (
351H351H
Fig. 3-13/
2
) on
the upper additional sleeves (
352H352H
Fig. 3-13/
1
) and
remove them by pulling them down as well.
•
Mount the additional sleeves again on the gear
knobs by proceeding in the reverse order.
Tighten the clamping screws when you are
done.
(5) Adjusting the friction (smoothness of operation) on both gear knobs of the ergonomics
The smoothness of ergonomics operation is factory-adjusted to a mid degree. The operator can change
the friction as follows:
a) How to adjust the X-direction
•
Remove the additional sleeves (
353H353H
Fig. 3-13/
1
and
3
) from the gear knobs by loosening the clamping
screws.
•
Shift the X-gear knob (
354H354H
Fig. 3-13/
5
) downward and the Y-gear knob (
355H355H
Fig. 3-13/
7
) upward.
•
Hold the X-gear knob (
356H356H
Fig. 3-13/
5
) and turn the light colored knurled ring above it (
357H357H
Fig. 3-13/
6
) to the
right (increased smoothness) or left (decreased smoothness) until you reach the desired degree.
b) How to adjust the Y-direction
•
Hold the Y-gear knob (
358H358H
Fig. 3-13/
7
) and turn the light colored knurled sleeve above it (
359H359H
Fig. 3-13/
8
) to
the right (increased smoothness) or the left (decreased smoothness) until you reach the desired
degree.
•
Replace the additional sleeves and tighten the clamping screws.
Enhance the life span of the stage by regularly removing the particles abraded from the object
carriers. Avoid getting any particles in the mechanical guiding system of the X-adjustment.
Fig. 3-13
Adjusting ergonometric drive
Summary of Contents for Axio Scope.A1
Page 1: ...Operating Manual Axio Scope A1 Microscope for Routine and Entry Level Research...
Page 23: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 23...
Page 24: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 24 M60 2 0007 e 05 08...
Page 25: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 25...
Page 26: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 26 M60 2 0007 e 05 08...
Page 122: ......
Page 123: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...