STARTUP
Carl Zeiss
Mounting the Standard Components
Axio Scope.A1
46
M60-2-0007 e 05/08
3.1.15
Halogen Lamp HAL 100
The halogen lamp HAL 100 may serve as light source for transmitted light and reflected light procedures
(except fluorescence light) as long as either the lower stand part or the upper stand part are equipped
with the appropriate connection piece. When mounting the halogen lamp, no difference is made
between reflected light and transmitted light procedures.
3.1.15.1
Mounting the halogen lamp
HAL 100
Before you switch on the halogen
lamp, please be sure to remove the
tools which are included in the lamp
case. The lamp generates a lot of heat
which could damage these tools (see
chapter
417H417H
3.1.15.3).
•
Remove the cap from the reflected light or
transmitted light connector.
•
Insert the lamp case (
418H418H
Fig. 3-23/
8
) with the
dovetail ring into the connector (
419H419H
Fig. 3-23/
2
and
420H420H
Fig. 3-23/
3
). Now fixate it by tightening the
clamping screw (
421H421H
Fig. 3-23/
1
or
422H422H
Fig. 3-23/
9
) with
the ball head screwdriver SW 3.
Due to the lack of ground clearance it
is important to avoid canting the lamp
when inserting it into the lower stand
part. Check for straight and correct fit.
•
Insert the 3-pin lamp plug (
423H423H
Fig. 3-23/
7
) into the
back of the external power supply (
424H424H
Fig. 3-23/
6
).
When mounting HAL 100 to the lower stand part for HAL 100 illumination, it is necessary to
use an additional control cord (
425H425H
Fig. 3-23/
5
) for the light intensity (see chapter
426H426H
3.2.2). The light
intensity of HAL 100 can then be adjusted with the control wheel (
427H427H
Fig. 3-23/
4
).
The light intensity of HAL 100 can also be adjusted with the external power supply unit. In this
case, the control cord needs to be removed.
Fig. 3-23
Mounting the halogen lamp
HAL 100
Summary of Contents for Axio Scope.A1
Page 1: ...Operating Manual Axio Scope A1 Microscope for Routine and Entry Level Research...
Page 23: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 23...
Page 24: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 24 M60 2 0007 e 05 08...
Page 25: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 25...
Page 26: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 26 M60 2 0007 e 05 08...
Page 122: ......
Page 123: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...