OPERATION
Axio Scope.A1
Operating and Functional Elements of Optional Components
Carl Zeiss
M60-2-0007 e 05/08
107
Mechanical stage 75x50
−
Mechanical stage (
717H717H
Fig. 4-34/
2
) with object
holder (
718H718H
Fig. 4-34/
1
) for holding, positioning
and fixating the samples.
−
Gear knobs for X (
719H719H
Fig. 4-34/
3
) and Y
adjustment (
720H720H
Fig. 4-34/
4
).
Overview fixture 2.5x-4x
The overview fixture serves to achieve a full
illumination of the field of view when using a
mildly magnifying objective (2.5x-4x) in
combination with the condenser 0.9/1.25
H
(424225-9000-000).
It can be centered and remains in the beam path
through the use of the according objective.
−
Swing the overview fixture (
721H721H
Fig. 4-35/
2
) in or
out of the beam path with the aid of the
handle (
722H722H
Fig. 4-35/
1
). Make sure to engage
the fixture securely after swinging it in.
−
Do not use it with LED illumination for
transmitted light.
An installation of the field diaphragm service unit is possible with the overhead fixture in place
if the condenser carrier is moved down to the lowest possible position!
Fig. 4-34
Mechanical stage with object
holder
Fig. 4-35
Overview fixture
Summary of Contents for Axio Scope.A1
Page 1: ...Operating Manual Axio Scope A1 Microscope for Routine and Entry Level Research...
Page 23: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 23...
Page 24: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 24 M60 2 0007 e 05 08...
Page 25: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 25...
Page 26: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 26 M60 2 0007 e 05 08...
Page 122: ......
Page 123: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...