STARTUP
Carl Zeiss
Mounting the Standard Components
Axio Scope.A1
40
M60-2-0007 e 05/08
(3) Adjusting the connecting rod position of
the ergo drive
The connecting rod position on the ergo drive of
the mechanical stage 75x50 R (432035-9010) can
be adjusted individually.
•
Remove the stage with the stage carrier from
the stand (see chapter
372H372H
3.1.13) and put it aside
face down.
•
Loosen both clamping screws (
373H373H
Fig. 3-15/
2
) on
the connecting rod (
374H374H
Fig. 3-15/
1
).
•
Slide the connecting rod (
375H375H
Fig. 3-15/
1
) in the
groove into the desired position.
•
Tighten both clamping screws (
376H376H
Fig. 3-15/
2
).
•
Mount the stage carrier with the stage on the
lower stand part (see chapter
377H377H
3.1.13).
Fig. 3-15
Adjusting the connecting rod
position for the ergo drive