INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
ZEISS
Disconnect the microscope from mains supply
Axio Observer
64
431004-7244-001
12/2016
4.20
Disconnect the microscope from mains supply
Axio Observer 3, 3 materials:
•
Disconnect the microscope from mains supply using the power switch on the microscope.
Axio Observer 5, 5 materials:
•
Shut down the internal computer of the microscope by using the standby button on the stand.
The device has not yet been disconnected from mains supply.
•
To disconnect the microscope unplug the power supply cable from the mains outlet socket.
Axio Observer 7, 7 materials:
•
Shut down the internal computer of the microscope by using the standby button on the stand.
The device has not yet been disconnected from mains supply.
•
To disconnect the microscope from the mains supply (e.g. If it is not used for a longer period) switch
off the power supply unit VP232-2 using its power switch.
4.21
Mounting the microLED illuminator
for transmitted and reflected light
The microLED illuminator can be used for both
transmitted light and reflected light.
Do not look directly into the LED light.
Always make sure that the instrument
is switched off when mounting the
microLED illuminator onto the
reflected light socket or dismounting
it.
•
Loosen clamping screw (Fig. 49/
1
) on reflected
light socket (Fig. 49/
2
) and remove the halogen
lamp.
•
Insert the attachment lamp (Fig. 49/
7
) with the
dovetail ring into the socket (Fig. 49/
2
) and,
using the 3 mm ball-headed screwdriver, fix it
using the clamping screw (Fig. 49/
1
).
•
Connect the 3-pin illuminator plug (Fig. 49/
6
) to the upper or lower 3-pin socket 12 V / 100 W for
transmitted light (Fig. 49/
3
) or (Fig. 49/
5
) for reflected light on the rear side of the instrument.
•
For the coded stands 3 and 5, set the transmitted light/reflected light toggle switch (Fig. 49/
4
) to the
corresponding position for transmitted light (DL) or reflected light (AL). If you use the motorized stand,
switching between reflected light and transmitted light is done via touch screen on the TFT display.
Fig. 49
Mounting the microLED illuminator